Misurazioni
direttiva
europea protezione UV
DIN
5050-1:2010-01 DIN
EN 60335-2-27 (VDE 0700-27):2009-04
Apparecchiatura
di misura per la prevenzione dai rischi derivanti dalle radiazioni
ottiche
artificiali non coerenti
X14
, Radiometro
di
precisione
per
misure UV su lettini abbronzanti
composto
da lettore e testa con 4 detectors.
European Commission UV
Protection
Directive
UV radiation is one of the most prevalent cancer risk factors. Because
of this,
federal lawmakers are trying to reduce the risk from artificial UV
radiation
with the UV Protection Directive. For sun tanning beds that emit UV
radiation,
this is achieved by a limit on the maximum UV output.
UV Erythema
Broadband Radiometer
In order to provide the responsible committees with a mobile instrument
to
measure UV radiation in accordance with the UV Protection Directive,
Gigahertz-Optik
GmbH has developed the X14
Broadband
Radiometer.
UVA
and UVB Two-channel UV Erythema
detector head
The UV radiometer‘s UV erythema
detector head sets
itself apart from others thanks to the fact that it features two UV
radiation
detectors. Separating the UVA and UVB spectral ranges between two
radiation
detectors reduces the crosstalk between the long-wavelength UVA and
short-wavelength UVB spectral ranges. This design principle offers a
greatly
reduced measurement uncertainty compared to UV erythema
measurement detectors with only one UV radiation detector.
Actinitic UV Erythema
Dose-Response Function
The spectral sensitivity of the two UV radiation detectors take
into account the UV erythema
actinic
function. Therefore
the sum signal of the two UV radiation detector forms the effective UV erythemal
irradiance. Additionally, both signals
can be
displayed separately in order to depict the ratio of UVB to UVA
radiation.
Detection
of UVC Radiation
In addition to the verification of the UV erythemal
irradiance, the UV Protection Directive conform inspection of UV sun
tanning
beds also requires verification of the UVC irradiance in the spectral
range
from 200 to 280nm. The UVC radiation must be less than the limit of
0.003 W/m².
The UV Eythema
measurement head of the X14
includes a separate UVC detector for this purpose.
UV
Protection Directive Conformity
The X14
Erythema
UV meter and
the
XD-45-ERYC measurement head were developed in accordance with DIN
60335-2-27
and DIN 5050-1 conform to the respective critical values.
Artificial
UV Sources and Sun Spectrum Correction
The measurement uncertainty due to the spectral mismatching of
Broadband
Radiometers can be effective reduced if the spectrum of the measured
lamp is
known. The X14 UV Erythema
Broadband
Radiometer
includes a database of common spectrums of artificial UV lamps,
technical UV
radiators, and the sun spectrum at various zenith angles. The
measurement
uncertainty of +/- 25% without knowledge of the lamp’s
spectrum
can be reduced
to +/- 6.5% when the spectrum is known.
Cosine-Corrected
Field of View
A
basic prerequisite for the accurate
measurement of
UV irradiance of extensive UV emitters is that the UV meter has a
cosine-corrected field of view. The UV Protection Directive refers to
the DIN
5050-1 that dictates an f2 error of less than 0.06.
Simple
Operation and Handling
Only
three buttons and a clearly structured
menu make
the meter easy to operate. The UV radiation detector head is separate
from the
device and is connected to the device using a 2m cable. There is a
mounting
point for a tripod on the measurement head.
Usage
with a Computer
The X14
UV meter includes a USB interface to enable
operation with a
computer. During USB operation the device is supplied with power
through the
USB interface.
Excel
Macro
To trigger a measurement and read out the result, the S-EXM-X1-4 Excel
macro is
optional available, which can be integrated into an Excel sheet by the
user.
X14
in the application to measure the UV-Erythema
and UVC
irradiation of sun tanning beds
Compact,
mobile UV lightmeter
Display of
the UV Erythema
effective and UVC
irradiance with
selected correction factor for solar radiation at 70 ° zenith
angle
X14
being used for the measurement of UV radiation due to
solar radiation.
XD-45-ERYC-4
UV erythema
detector head with additional
UVC
radiation detection function.
Relevant norms
and regulations |
UV Protection
Directive |
Threshold values
irradiance |
|
UV
Protection
Directive |
250-400nm 0,3
W/m²eff |
60335-2-27 |
Identical to UV
Protection Directive, except for additional subdivisions for the
wavelength |
5050-1 |
No threshold
value defined for UVC 200-290nm |
Threshold values
spectral
correction |
|
UV
Protection
Directive |
Consult
60335-2-27
and 5050-1 |
60335-2-27 |
No value for
spectral correction |
5050-1 |
f1<
0.2 for broadband radiometer without specification of reference spectrum |
Threshold value
cosine correction |
|
UV
Protection
Directive |
Consult
60335-2-27and
5050-1 |
60335-2-27 |
No
specifications for difference to cosine |
5050-1 |
f2
<0.06 for
broadband
radiometer |
Threshold value
temperature coefficient |
|
UV
Protection
Directive |
Consult
60335-2-27and
5050-1 |
60335-2-27 |
Measurement at
20±5°C ambient temperature |
5050-1 |
0,005
K-1
(10-50°C) |
Burn-in period of UV Emitters |
|
UV
Protection
Directive |
Consult
60335-2-27and
5050-1 |
60335-2-27 |
Half of the
maximum exposure time |
5050-1 |
At least 15min
and fluctuations less than ±5% |
Exposure time for 100J/m² |
|
UV Protection
Directive calculation example |
At Eeff
0.3W/m² about 5min und 30s |
Threshold value
measurement
distance |
|
UV
Protection
Directive |
Consult
60335-2-27and
5050-1 |
60335-2-27 |
Min. recommended
distance 100±2mm for facial radiation devices |
5050-1 |
Full body
radiation 0mm or in a radius of 30cm |