Misurazioni direttiva europea protezione UV   DIN 5050-1:2010-01  DIN EN 60335-2-27 (VDE 0700-27):2009-04

Apparecchiatura di misura per la prevenzione dai rischi derivanti dalle radiazioni ottiche artificiali non coerenti

 

X14, UV Erythema Broadband Radiometer for Hazard Testing Sun Tanning Beds

 

X14  , Radiometro di precisione
 per misure UV su lettini abbronzanti

composto da lettore e testa con 4 detectors.

 

European Commission UV Protection Directive
UV radiation is one of the most prevalent cancer risk factors. Because of this, federal lawmakers are trying to reduce the risk from artificial UV radiation with the UV Protection Directive. For sun tanning beds that emit UV radiation, this is achieved by a limit on the maximum UV output.

UV Erythema Broadband Radiometer
In order to provide the responsible committees with a mobile instrument to measure UV radiation in accordance with the UV Protection Directive, Gigahertz-Optik GmbH has developed the X14 Broadband Radiometer.

UVA and UVB Two-channel UV Erythema detector head
The UV radiometer‘s UV erythema detector head sets itself apart from others thanks to the fact that it features two UV radiation detectors. Separating the UVA and UVB spectral ranges between two radiation detectors reduces the crosstalk between the long-wavelength UVA and short-wavelength UVB spectral ranges. This design principle offers a greatly reduced measurement uncertainty compared to UV erythema measurement detectors with only one UV radiation detector.

Actinitic UV Erythema Dose-Response Function
The spectral sensitivity of the two UV radiation detectors take into account the UV erythema actinic function. Therefore the sum signal of the two UV radiation detector forms the effective UV erythemal irradiance. Additionally, both signals can be displayed separately in order to depict the ratio of UVB to UVA radiation.

Detection of UVC Radiation
In addition to the verification of the UV erythemal irradiance, the UV Protection Directive conform inspection of UV sun tanning beds also requires verification of the UVC irradiance in the spectral range from 200 to 280nm. The UVC radiation must be less than the limit of 0.003 W/m². The UV Eythema measurement head of the X14 includes a separate UVC detector for this purpose.

UV Protection Directive Conformity
The X14 Erythema UV meter and the XD-45-ERYC measurement head were developed in accordance with DIN 60335-2-27 and DIN 5050-1 conform to the respective critical values.

Artificial UV Sources and Sun Spectrum Correction
The measurement uncertainty due to the spectral mismatching of Broadband Radiometers can be effective reduced if the spectrum of the measured lamp is known. The X14 UV Erythema Broadband Radiometer includes a database of common spectrums of artificial UV lamps, technical UV radiators, and the sun spectrum at various zenith angles. The measurement uncertainty of +/- 25% without knowledge of the lamp’s spectrum can be reduced to +/- 6.5% when the spectrum is known.

Cosine-Corrected Field of View
A basic prerequisite for the accurate measurement of UV irradiance of extensive UV emitters is that the UV meter has a cosine-corrected field of view. The UV Protection Directive refers to the DIN 5050-1 that dictates an f2 error of less than 0.06.

Simple Operation and Handling
Only three buttons and a clearly structured menu make the meter easy to operate. The UV radiation detector head is separate from the device and is connected to the device using a 2m cable. There is a mounting point for a tripod on the measurement head.

Usage with a Computer
The X14 UV meter includes a USB interface to enable operation with a computer. During USB operation the device is supplied with power through the USB interface.

Excel Macro
To trigger a measurement and read out the result, the S-EXM-X1-4 Excel macro is optional available, which can be integrated into an Excel sheet by the user.

X14 in the application to measure the UV-Erythema and UVC irradiation of sun tanning beds



Compact, mobile UV lightmeter

 

Display of the UV Erythema effective and UVC irradiance with selected correction factor for solar radiation at 70 ° zenith angle

 

X14 being used for the measurement of UV radiation due to solar radiation.

 

XD-45-ERYC-4 UV erythema detector head with additional UVC radiation detection function.

 

Relevant norms and regulations

UV Protection Directive
DIN EN 60335-2-27 (VDE 0700-27):2009-04
DIN 5050-1:2010-01

Threshold values irradiance

 

UV Protection Directive

250-400nm 0,3 W/m²eff
200-280nm 0,003 W/m²

60335-2-27

Identical to UV Protection Directive, except for additional subdivisions for the wavelength
250-320nm and 320-400nm

5050-1

No threshold value defined for UVC 200-290nm

Threshold values spectral correction

 

UV Protection Directive

Consult 60335-2-27 and 5050-1

60335-2-27

No value for spectral correction

5050-1

f1< 0.2 for broadband radiometer without specification of reference spectrum

Threshold value cosine correction

 

UV Protection Directive

Consult 60335-2-27and 5050-1

60335-2-27

No specifications for difference to cosine

5050-1

f2 <0.06 for broadband radiometer

Threshold value temperature coefficient

 

UV Protection Directive

Consult 60335-2-27and 5050-1

60335-2-27

Measurement at 20±5°C ambient temperature

5050-1

0,005 K-1 (10-50°C)

Burn-in period of UV Emitters

 

UV Protection Directive

Consult 60335-2-27and 5050-1

60335-2-27

Half of the maximum exposure time

5050-1

At least 15min and fluctuations less than ±5%

Exposure time for 100J/m²

 

UV Protection Directive calculation example

At Eeff 0.3W/m² about 5min und 30s
100J/m² / 0.3W/m² ≈ 330s

Threshold value measurement distance

 

UV Protection Directive

Consult 60335-2-27and 5050-1

60335-2-27

Min. recommended distance 100±2mm for facial radiation devices

5050-1

Full body radiation 0mm or in a radius of 30cm
in front of the measurement point on the support surface

 

 

 

 