AvaSpec-3648 Fiber Optic
Spectrometer - Spettrometro 3648 canali
Technical Details
Optical Bench |
Symmetrical Czerny-Turner, 75 mm focal length |
Wavelength range |
200 - 1100nm |
Resolution |
0.025 –20 nm, depending on configuration (see table) |
Stray light |
< 0.1% |
Sensitivity (AvaLight-HAL, 8 µm fiber) |
12000 counts/µW per ms integration time |
Detector |
CCD linear array, 3648 pixels |
Signal/Noise |
300:1 |
AD converter |
14 bit, 1 MHz |
Integration time |
10 µsec – 10 min. |
Interface |
USB 2.0 high speed, 480 Mbps |
Power supply |
Default USB power, 440 mA |
Dimensions, weight |
175 x 110 x 44 mm(1 channel), 716 grams |
Grating selection table for AvaSpec-3648
Use |
Useable range |
Spectral range (nm) |
Lines/mm |
Blaze (nm) |
Order code |
UV/VIS/NIR |
200-1100** |
900** |
300 |
300 |
UA |
UV/VIS |
200-850 |
520 |
600 |
250 |
UB |
UV |
200-750 |
250-220* |
1200 |
250 |
UC |
UV |
200-650 |
165-145* |
1800 |
250 |
UD |
UV |
200-580 |
115-70* |
2400 |
250 |
UE |
UV |
220-400 |
75-50* |
3600 |
250 |
UF |
UV/VIS |
250-850 |
520 |
600 |
370 |
BB |
VIS/NIR |
300-1100** |
800** |
300 |
500 |
VA |
VIS |
360-1000 |
500 |
600 |
500 |
VB |
VIS |
300-800 |
250-200* |
1200 |
500 |
VC |
VIS |
350-750 |
145-100* |
1800 |
500 |
VD |
NIR |
500-1050 |
500 |
600 |
750 |
NB |
NIR |
500-1050 |
220-150* |
1200 |
750 |
NC |
NIR |
600-1100** |
500** |
300 |
1000 |
IA |
NIR |
600-1100 |
500 |
600 |
1000 |
IB |
* depends on the starting
wavelength of the grating; the higher the wavelength, the bigger the dispersion
and the smaller the range
** please note that all 2048 pixels will be used for the useable range
Resolution table (FWHM) for AvaSpec-3648
Slit size (µm) |
||||||
Grating (lines/mm) |
10 |
25 |
50 |
100 |
200 |
500 |
300 |
0.5 |
1.4 |
2.4 |
4.3 |
8.0 |
20.0 |
600 |
0.25 |
0.7 |
1.2 |
2.1 |
4.1 |
10.0 |
1200 |
0.07-0.13* |
0.2-0.3* |
0.4-0.6* |
0.7-1.0* |
1.4-2.0* |
3.3-4.8* |
1800 |
0.05-0.08* |
0.12-0.21* |
0.2-0.36* |
0.4-0.7* |
0.7-1.4* |
1.7-3.3* |
2400 |
0.04-0.07* |
0.08-0.15* |
0.14-0.25* |
0.3-0.5* |
0.5-0.9* |
1.2-2.2* |
3600 |
0.025-0.04* |
0.07-0.10* |
0.11-0.16* |
0.2-0.3* |
0.4-0.6* |
0.9-1.4* |
* = depends on the starting wavelength of the grating; the higher the wavelength, the bigger the dispersion and the better the resolution
Home | Prodotti | Lista rappresentate | Modulo Informazioni | News | Contatti